Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom

Descripción del Articulo

In radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of...

Descripción completa

Detalles Bibliográficos
Autores: Querebalú, Cecilia, Solano, Javier, Márquez, Fernando
Formato: artículo
Fecha de Publicación:2018
Institución:Universidad Nacional de Ingeniería
Repositorio:Revistas - Universidad Nacional de Ingeniería
Lenguaje:español
OAI Identifier:oai:oai:revistas.uni.edu.pe:article/343
Enlace del recurso:https://revistas.uni.edu.pe/index.php/tecnia/article/view/343
Nivel de acceso:acceso abierto
Materia:radiocirugía
dosis periférica
películas radiocrómicas EBT2
radiosurgery
peripheral dose
EBT2 radiochromic films
id REVUNI_2b60ad728af9d67ba137fa76367e2499
oai_identifier_str oai:oai:revistas.uni.edu.pe:article/343
network_acronym_str REVUNI
network_name_str Revistas - Universidad Nacional de Ingeniería
repository_id_str
spelling Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantomEvaluación de la dosis periférica para tratamientos de radiocirugía en blancos simulados en diferentes posiciones de la cabeza de un fantomaQuerebalú, CeciliaSolano, JavierMárquez, Fernandoradiocirugíadosis periféricapelículas radiocrómicas EBT2radiosurgeryperipheral doseEBT2 radiochromic filmsIn radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of an anthropomorphic phantom. The development of the work contemplates the use of the EBT2 radiochromic films, and a photon beam of 6 MV generated in a linear accelerator Clinac 2100C / D. Analyzing the influence of collimator size, greater DP ((150.81 ± 0.97) cGy) was found in the right eye of the phantom for the collimator of 3 cm with a target fixed in the upper right plane of the head of the phantom and its influence with the distance of, a PD of (27.14 ± 1.04) cGy in the thyroid, for a target positioned in the lower left plane of the phantom head for a prescribed dose of 18 Gy. It is concluded that to carry out the optimization during the procedure in radiosurgery it is necessary to know the dependence of the PD with respect to the location of the target, the size of collimator and the distance. En tratamientos de radiocirugía siempre hay una dosis absorbida (D) que se deposita fuera del campo de tratamiento, a esta D se le conoce como “Dosis Periférica” (DP). El objetivo de este trabajo fue la evaluación de DP en 9 posiciones anatómicas (ojos, tiroides, parte superior de los pulmones, mediastino, mamas y ombligo) de un fantoma antropomórfico. El desarrollo del trabajo contemplo el uso de películas radiocrómicas EBT2, y un haz de fotones de 6 MV generados en un acelerador lineal Clinac 2100C/D. Analizando la influencia del tamaño de colimador, se encontró mayor DP ((150.81 ± 0.97) cGy) en el ojo derecho del fantoma para el  colimador de 3 cm con un blanco posicionado en el plano superior derecho de la cabeza del fantoma y su influencia con la distancia de (27.14 ± 1.04) cGy en la tiroides, para un blanco posicionado en el plano inferior izquierda de la cabeza del fantoma para una dosis prescrita de 18 Gy. Se concluye que para realizar la optimización durante el procedimiento en radiocirugía se tiene que conocer la dependencia de la DP con respecto a la localización del blanco, el tamaño de colimador y la distancia.Universidad Nacional de Ingeniería2018-12-17info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionMedical physicsFísica Médicaapplication/pdfaudio/mpegtext/plainhttps://revistas.uni.edu.pe/index.php/tecnia/article/view/34310.21754/tecnia.v28i2.343TECNIA; Vol. 28 No. 2 (2018)TECNIA; Vol. 28 Núm. 2 (2018)2309-04130375-7765reponame:Revistas - Universidad Nacional de Ingenieríainstname:Universidad Nacional de Ingenieríainstacron:UNIspahttps://revistas.uni.edu.pe/index.php/tecnia/article/view/343/575https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/768https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/977Derechos de autor 2018 TECNIAhttp://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessoai:oai:revistas.uni.edu.pe:article/3432023-12-06T21:58:37Z
dc.title.none.fl_str_mv Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
Evaluación de la dosis periférica para tratamientos de radiocirugía en blancos simulados en diferentes posiciones de la cabeza de un fantoma
title Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
spellingShingle Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
Querebalú, Cecilia
radiocirugía
dosis periférica
películas radiocrómicas EBT2
radiosurgery
peripheral dose
EBT2 radiochromic films
title_short Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
title_full Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
title_fullStr Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
title_full_unstemmed Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
title_sort Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
dc.creator.none.fl_str_mv Querebalú, Cecilia
Solano, Javier
Márquez, Fernando
author Querebalú, Cecilia
author_facet Querebalú, Cecilia
Solano, Javier
Márquez, Fernando
author_role author
author2 Solano, Javier
Márquez, Fernando
author2_role author
author
dc.subject.none.fl_str_mv radiocirugía
dosis periférica
películas radiocrómicas EBT2
radiosurgery
peripheral dose
EBT2 radiochromic films
topic radiocirugía
dosis periférica
películas radiocrómicas EBT2
radiosurgery
peripheral dose
EBT2 radiochromic films
description In radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of an anthropomorphic phantom. The development of the work contemplates the use of the EBT2 radiochromic films, and a photon beam of 6 MV generated in a linear accelerator Clinac 2100C / D. Analyzing the influence of collimator size, greater DP ((150.81 ± 0.97) cGy) was found in the right eye of the phantom for the collimator of 3 cm with a target fixed in the upper right plane of the head of the phantom and its influence with the distance of, a PD of (27.14 ± 1.04) cGy in the thyroid, for a target positioned in the lower left plane of the phantom head for a prescribed dose of 18 Gy. It is concluded that to carry out the optimization during the procedure in radiosurgery it is necessary to know the dependence of the PD with respect to the location of the target, the size of collimator and the distance. 
publishDate 2018
dc.date.none.fl_str_mv 2018-12-17
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Medical physics
Física Médica
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv https://revistas.uni.edu.pe/index.php/tecnia/article/view/343
10.21754/tecnia.v28i2.343
url https://revistas.uni.edu.pe/index.php/tecnia/article/view/343
identifier_str_mv 10.21754/tecnia.v28i2.343
dc.language.none.fl_str_mv spa
language spa
dc.relation.none.fl_str_mv https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/575
https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/768
https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/977
dc.rights.none.fl_str_mv Derechos de autor 2018 TECNIA
http://creativecommons.org/licenses/by/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Derechos de autor 2018 TECNIA
http://creativecommons.org/licenses/by/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
audio/mpeg
text/plain
dc.publisher.none.fl_str_mv Universidad Nacional de Ingeniería
publisher.none.fl_str_mv Universidad Nacional de Ingeniería
dc.source.none.fl_str_mv TECNIA; Vol. 28 No. 2 (2018)
TECNIA; Vol. 28 Núm. 2 (2018)
2309-0413
0375-7765
reponame:Revistas - Universidad Nacional de Ingeniería
instname:Universidad Nacional de Ingeniería
instacron:UNI
instname_str Universidad Nacional de Ingeniería
instacron_str UNI
institution UNI
reponame_str Revistas - Universidad Nacional de Ingeniería
collection Revistas - Universidad Nacional de Ingeniería
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1833562776988024832
score 13.982353
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).