Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom
Descripción del Articulo
In radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of...
| Autores: | , , |
|---|---|
| Formato: | artículo |
| Fecha de Publicación: | 2018 |
| Institución: | Universidad Nacional de Ingeniería |
| Repositorio: | Revistas - Universidad Nacional de Ingeniería |
| Lenguaje: | español |
| OAI Identifier: | oai:oai:revistas.uni.edu.pe:article/343 |
| Enlace del recurso: | https://revistas.uni.edu.pe/index.php/tecnia/article/view/343 |
| Nivel de acceso: | acceso abierto |
| Materia: | radiocirugía dosis periférica películas radiocrómicas EBT2 radiosurgery peripheral dose EBT2 radiochromic films |
| id |
REVUNI_2b60ad728af9d67ba137fa76367e2499 |
|---|---|
| oai_identifier_str |
oai:oai:revistas.uni.edu.pe:article/343 |
| network_acronym_str |
REVUNI |
| network_name_str |
Revistas - Universidad Nacional de Ingeniería |
| repository_id_str |
|
| spelling |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantomEvaluación de la dosis periférica para tratamientos de radiocirugía en blancos simulados en diferentes posiciones de la cabeza de un fantomaQuerebalú, CeciliaSolano, JavierMárquez, Fernandoradiocirugíadosis periféricapelículas radiocrómicas EBT2radiosurgeryperipheral doseEBT2 radiochromic filmsIn radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of an anthropomorphic phantom. The development of the work contemplates the use of the EBT2 radiochromic films, and a photon beam of 6 MV generated in a linear accelerator Clinac 2100C / D. Analyzing the influence of collimator size, greater DP ((150.81 ± 0.97) cGy) was found in the right eye of the phantom for the collimator of 3 cm with a target fixed in the upper right plane of the head of the phantom and its influence with the distance of, a PD of (27.14 ± 1.04) cGy in the thyroid, for a target positioned in the lower left plane of the phantom head for a prescribed dose of 18 Gy. It is concluded that to carry out the optimization during the procedure in radiosurgery it is necessary to know the dependence of the PD with respect to the location of the target, the size of collimator and the distance. En tratamientos de radiocirugía siempre hay una dosis absorbida (D) que se deposita fuera del campo de tratamiento, a esta D se le conoce como “Dosis Periférica” (DP). El objetivo de este trabajo fue la evaluación de DP en 9 posiciones anatómicas (ojos, tiroides, parte superior de los pulmones, mediastino, mamas y ombligo) de un fantoma antropomórfico. El desarrollo del trabajo contemplo el uso de películas radiocrómicas EBT2, y un haz de fotones de 6 MV generados en un acelerador lineal Clinac 2100C/D. Analizando la influencia del tamaño de colimador, se encontró mayor DP ((150.81 ± 0.97) cGy) en el ojo derecho del fantoma para el colimador de 3 cm con un blanco posicionado en el plano superior derecho de la cabeza del fantoma y su influencia con la distancia de (27.14 ± 1.04) cGy en la tiroides, para un blanco posicionado en el plano inferior izquierda de la cabeza del fantoma para una dosis prescrita de 18 Gy. Se concluye que para realizar la optimización durante el procedimiento en radiocirugía se tiene que conocer la dependencia de la DP con respecto a la localización del blanco, el tamaño de colimador y la distancia.Universidad Nacional de Ingeniería2018-12-17info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionMedical physicsFísica Médicaapplication/pdfaudio/mpegtext/plainhttps://revistas.uni.edu.pe/index.php/tecnia/article/view/34310.21754/tecnia.v28i2.343TECNIA; Vol. 28 No. 2 (2018)TECNIA; Vol. 28 Núm. 2 (2018)2309-04130375-7765reponame:Revistas - Universidad Nacional de Ingenieríainstname:Universidad Nacional de Ingenieríainstacron:UNIspahttps://revistas.uni.edu.pe/index.php/tecnia/article/view/343/575https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/768https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/977Derechos de autor 2018 TECNIAhttp://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessoai:oai:revistas.uni.edu.pe:article/3432023-12-06T21:58:37Z |
| dc.title.none.fl_str_mv |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom Evaluación de la dosis periférica para tratamientos de radiocirugía en blancos simulados en diferentes posiciones de la cabeza de un fantoma |
| title |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| spellingShingle |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom Querebalú, Cecilia radiocirugía dosis periférica películas radiocrómicas EBT2 radiosurgery peripheral dose EBT2 radiochromic films |
| title_short |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| title_full |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| title_fullStr |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| title_full_unstemmed |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| title_sort |
Evaluation of the peripheral dose for radiosurgery treatments in simulated targets in different positions of the head of a phantom |
| dc.creator.none.fl_str_mv |
Querebalú, Cecilia Solano, Javier Márquez, Fernando |
| author |
Querebalú, Cecilia |
| author_facet |
Querebalú, Cecilia Solano, Javier Márquez, Fernando |
| author_role |
author |
| author2 |
Solano, Javier Márquez, Fernando |
| author2_role |
author author |
| dc.subject.none.fl_str_mv |
radiocirugía dosis periférica películas radiocrómicas EBT2 radiosurgery peripheral dose EBT2 radiochromic films |
| topic |
radiocirugía dosis periférica películas radiocrómicas EBT2 radiosurgery peripheral dose EBT2 radiochromic films |
| description |
In radiosurgery treatments, there is always an absorbed dose (D) that is deposited outside the treatment field, this D is known as "Peripheral Dose" (PD). The objective of this work was to evaluate PD in 9 anatomical positions (eyes, thyroid, upper lungs, mediastinum, breasts and navel) of an anthropomorphic phantom. The development of the work contemplates the use of the EBT2 radiochromic films, and a photon beam of 6 MV generated in a linear accelerator Clinac 2100C / D. Analyzing the influence of collimator size, greater DP ((150.81 ± 0.97) cGy) was found in the right eye of the phantom for the collimator of 3 cm with a target fixed in the upper right plane of the head of the phantom and its influence with the distance of, a PD of (27.14 ± 1.04) cGy in the thyroid, for a target positioned in the lower left plane of the phantom head for a prescribed dose of 18 Gy. It is concluded that to carry out the optimization during the procedure in radiosurgery it is necessary to know the dependence of the PD with respect to the location of the target, the size of collimator and the distance. |
| publishDate |
2018 |
| dc.date.none.fl_str_mv |
2018-12-17 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion Medical physics Física Médica |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
https://revistas.uni.edu.pe/index.php/tecnia/article/view/343 10.21754/tecnia.v28i2.343 |
| url |
https://revistas.uni.edu.pe/index.php/tecnia/article/view/343 |
| identifier_str_mv |
10.21754/tecnia.v28i2.343 |
| dc.language.none.fl_str_mv |
spa |
| language |
spa |
| dc.relation.none.fl_str_mv |
https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/575 https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/768 https://revistas.uni.edu.pe/index.php/tecnia/article/view/343/977 |
| dc.rights.none.fl_str_mv |
Derechos de autor 2018 TECNIA http://creativecommons.org/licenses/by/4.0 info:eu-repo/semantics/openAccess |
| rights_invalid_str_mv |
Derechos de autor 2018 TECNIA http://creativecommons.org/licenses/by/4.0 |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf audio/mpeg text/plain |
| dc.publisher.none.fl_str_mv |
Universidad Nacional de Ingeniería |
| publisher.none.fl_str_mv |
Universidad Nacional de Ingeniería |
| dc.source.none.fl_str_mv |
TECNIA; Vol. 28 No. 2 (2018) TECNIA; Vol. 28 Núm. 2 (2018) 2309-0413 0375-7765 reponame:Revistas - Universidad Nacional de Ingeniería instname:Universidad Nacional de Ingeniería instacron:UNI |
| instname_str |
Universidad Nacional de Ingeniería |
| instacron_str |
UNI |
| institution |
UNI |
| reponame_str |
Revistas - Universidad Nacional de Ingeniería |
| collection |
Revistas - Universidad Nacional de Ingeniería |
| repository.name.fl_str_mv |
|
| repository.mail.fl_str_mv |
|
| _version_ |
1833562776988024832 |
| score |
13.982353 |
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).