VALIDATION OF A METHOD BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY FOR THE QUANTIFICATION OF SENNOSIDES IN CHOCOLATE BARS

Descripción del Articulo

To develop a new product pharmaceutical industries, use analytical methods to quantify the active principle fast and reliable. The validation was determined if the method is reliable and reproducible, for this reason the present work validated an analytical method by high resolution liquid chromatog...

Descripción completa

Detalles Bibliográficos
Autores: Samaniego, Jhonnel, Huerta, Jenny, Inocente, Miguel, Obregón, Juan, López, Maricela
Formato: artículo
Fecha de Publicación:2021
Institución:Sociedad Química del Perú
Repositorio:Revista de la Sociedad Química del Perú
Lenguaje:español
OAI Identifier:oai:rsqp.revistas.sqperu.org.pe:article/338
Enlace del recurso:http://revistas.sqperu.org.pe/index.php/revistasqperu/article/view/338
Nivel de acceso:acceso abierto
Materia:validation
high performance liquid chromatography
sennosides
validación
cromatografía líquida de alta resolución
senósidos
Descripción
Sumario:To develop a new product pharmaceutical industries, use analytical methods to quantify the active principle fast and reliable. The validation was determined if the method is reliable and reproducible, for this reason the present work validated an analytical method by high resolution liquid chromatography to quantify sennosides in chocolate bars, the type of research was descriptive with simple random sampling. The chromatographic conditions at 274 nm, mobile phase with a flow gradient ranging from 0.4 to 1, 5 mL / minute and solvent, composed to phosphate buffer pH 3.5 and Acetonitrile, injection volume 40 μL. The evaluated validation parameters confirmed specificity, linearity to concentrations of 0.07; 0.11; 0.15; 0.19 and 0.23 mg / mL with a correlation coefficient of 0.9996, precision with a coefficient of variation less than 2 %, accuracy of 98.6 % and robustness after 24 hours with C.V. (%) ≤ 2 %. The proposed method that the proposed method is suitable for the quantification of sennosides in chocolate bars.
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).