Characterizing a CCD detector for astronomical purposes: OAUNI Project
Descripción del Articulo
This work verifies the instrumental characteristics of the CCD detector which is part of the UNI astronomical observatory. We measured the linearity of the CCD detector of the SBIG STXL6303E camera, along with the associated gain and readout noise. The linear response to the incident light of the de...
| Autores: | , , , |
|---|---|
| Formato: | artículo |
| Fecha de Publicación: | 2016 |
| Institución: | Universidad Nacional de Ingeniería |
| Repositorio: | Revista UNI - Tecnia |
| Lenguaje: | español |
| OAI Identifier: | oai:oai:revistas.uni.edu.pe:article/54 |
| Enlace del recurso: | http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54 |
| Nivel de acceso: | acceso abierto |
| Materia: | Dispositivos de carga acoplada Linealidad ganancia ruido de lectura |
| id |
0375-7765_eec35b0de78f02d8847d5ac769baba23 |
|---|---|
| oai_identifier_str |
oai:oai:revistas.uni.edu.pe:article/54 |
| network_acronym_str |
0375-7765 |
| repository_id_str |
. |
| network_name_str |
Revista UNI - Tecnia |
| spelling |
Characterizing a CCD detector for astronomical purposes: OAUNI ProjectCaracterizando un detector CCD para uso astronómico: proyecto OAUNIPereyra, AntonioZevallos, María I.Ricra, JoséTello, Julio C.Dispositivos de carga acopladaLinealidadgananciaruido de lecturaThis work verifies the instrumental characteristics of the CCD detector which is part of the UNI astronomical observatory. We measured the linearity of the CCD detector of the SBIG STXL6303E camera, along with the associated gain and readout noise. The linear response to the incident light of the detector is extremely linear (R2 =99.99%), its effective gain is 1.65 ± 0.01 e-/ADU and its readout noise is 12.2 e-. These values are in agreement with the manufacturer. We confirm that this detector is extremely precise to make measurements for astronomical purposes. El presente trabajo verificó las capacidades del detector CCD que forma parte del instrumental asociado al proyecto del Observatorio Astronómico de la UNI. Fueron medidos la respuesta lineal del detector CCD de la cámara SBIG STXL6303E, así como su ganancia y el ruido de lectura asociados. La respuesta lineal a la luz incidente del detector es extremadamente lineal (R2=99.99%), su ganancia efectiva es 1.65 ± 0.1 e-/ADU y su ruido de lectura 12.2 e-. Estos valores están de acuerdo con los proporcionados por el fabricante. Se confirma que el detector analizado es extremadamente preciso para realizar medidas con fines astronómicos. Universidad Nacional de Ingeniería2016-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionArtículo evaluado por paresapplication/pdfaudio/mpegtext/htmlhttp://www.revistas.uni.edu.pe/index.php/tecnia/article/view/5410.21754/tecnia.v26i2.54TECNIA; Vol 26 No 2 (2016); 20TECNIA; Vol. 26 Núm. 2 (2016); 202309-04130375-7765reponame:Revista UNI - Tecniainstname:Universidad Nacional de Ingenieríainstacron:UNIspahttp://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/100http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/500http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/561info:eu-repo/semantics/openAccess2021-05-29T15:55:17Zmail@mail.com - |
| dc.title.none.fl_str_mv |
Characterizing a CCD detector for astronomical purposes: OAUNI Project Caracterizando un detector CCD para uso astronómico: proyecto OAUNI |
| title |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| spellingShingle |
Characterizing a CCD detector for astronomical purposes: OAUNI Project Pereyra, Antonio Dispositivos de carga acoplada Linealidad ganancia ruido de lectura |
| title_short |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| title_full |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| title_fullStr |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| title_full_unstemmed |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| title_sort |
Characterizing a CCD detector for astronomical purposes: OAUNI Project |
| dc.creator.none.fl_str_mv |
Pereyra, Antonio Zevallos, María I. Ricra, José Tello, Julio C. |
| author |
Pereyra, Antonio |
| author_facet |
Pereyra, Antonio Zevallos, María I. Ricra, José Tello, Julio C. |
| author_role |
author |
| author2 |
Zevallos, María I. Ricra, José Tello, Julio C. |
| author2_role |
author author author |
| dc.subject.none.fl_str_mv |
Dispositivos de carga acoplada Linealidad ganancia ruido de lectura |
| topic |
Dispositivos de carga acoplada Linealidad ganancia ruido de lectura |
| dc.description.none.fl_txt_mv |
This work verifies the instrumental characteristics of the CCD detector which is part of the UNI astronomical observatory. We measured the linearity of the CCD detector of the SBIG STXL6303E camera, along with the associated gain and readout noise. The linear response to the incident light of the detector is extremely linear (R2 =99.99%), its effective gain is 1.65 ± 0.01 e-/ADU and its readout noise is 12.2 e-. These values are in agreement with the manufacturer. We confirm that this detector is extremely precise to make measurements for astronomical purposes. El presente trabajo verificó las capacidades del detector CCD que forma parte del instrumental asociado al proyecto del Observatorio Astronómico de la UNI. Fueron medidos la respuesta lineal del detector CCD de la cámara SBIG STXL6303E, así como su ganancia y el ruido de lectura asociados. La respuesta lineal a la luz incidente del detector es extremadamente lineal (R2=99.99%), su ganancia efectiva es 1.65 ± 0.1 e-/ADU y su ruido de lectura 12.2 e-. Estos valores están de acuerdo con los proporcionados por el fabricante. Se confirma que el detector analizado es extremadamente preciso para realizar medidas con fines astronómicos. |
| description |
This work verifies the instrumental characteristics of the CCD detector which is part of the UNI astronomical observatory. We measured the linearity of the CCD detector of the SBIG STXL6303E camera, along with the associated gain and readout noise. The linear response to the incident light of the detector is extremely linear (R2 =99.99%), its effective gain is 1.65 ± 0.01 e-/ADU and its readout noise is 12.2 e-. These values are in agreement with the manufacturer. We confirm that this detector is extremely precise to make measurements for astronomical purposes. |
| publishDate |
2016 |
| dc.date.none.fl_str_mv |
2016-12-01 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion Artículo evaluado por pares |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54 10.21754/tecnia.v26i2.54 |
| url |
http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54 |
| identifier_str_mv |
10.21754/tecnia.v26i2.54 |
| dc.language.none.fl_str_mv |
spa |
| language |
spa |
| dc.relation.none.fl_str_mv |
http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/100 http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/500 http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/54/561 |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf audio/mpeg text/html |
| dc.publisher.none.fl_str_mv |
Universidad Nacional de Ingeniería |
| publisher.none.fl_str_mv |
Universidad Nacional de Ingeniería |
| dc.source.none.fl_str_mv |
TECNIA; Vol 26 No 2 (2016); 20 TECNIA; Vol. 26 Núm. 2 (2016); 20 2309-0413 0375-7765 reponame:Revista UNI - Tecnia instname:Universidad Nacional de Ingeniería instacron:UNI |
| reponame_str |
Revista UNI - Tecnia |
| collection |
Revista UNI - Tecnia |
| instname_str |
Universidad Nacional de Ingeniería |
| instacron_str |
UNI |
| institution |
UNI |
| repository.name.fl_str_mv |
-
|
| repository.mail.fl_str_mv |
mail@mail.com |
| _version_ |
1701108798898307072 |
| score |
13.924177 |
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).