Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters

Descripción del Articulo

The choice of the attenuation method to characterize the X rays beams used in the calibration of personal dosemeters at the Calibraion and Dosimetry laboratiry of the Peruvian Institute of Nuclear Energy (IPEN). Voltages applied to the X ray tube and the addition of filters have been made according...

Descripción completa

Detalles Bibliográficos
Autores: Benavente, T., Márquez, J., Santa Cruz, J.
Formato: artículo
Fecha de Publicación:1999
Institución:Universidad Nacional Mayor de San Marcos
Repositorio:Revistas - Universidad Nacional Mayor de San Marcos
Lenguaje:español
OAI Identifier:oai:ojs.csi.unmsm:article/8599
Enlace del recurso:https://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/8599
Nivel de acceso:acceso abierto
Materia:X ray
attenuation
spectra
dosimetry
radioprotection
rayos X
atenuación
espectros
dosímetría
radioprotección
id REVUNMSM_34348670e97c0938505dd041dabc57e2
oai_identifier_str oai:ojs.csi.unmsm:article/8599
network_acronym_str REVUNMSM
network_name_str Revistas - Universidad Nacional Mayor de San Marcos
repository_id_str
spelling Characterization of X-ray beams by Attenuation Method for Calibration of Personal DosimetersCaracterización de Haces de Rayos X por el Método de Atenuación para la Calibración de Dosímetros PersonalesBenavente, T.Márquez, J.Santa Cruz, J.X rayattenuationspectradosimetryradioprotectionrayos XatenuaciónespectrosdosímetríaradioprotecciónThe choice of the attenuation method to characterize the X rays beams used in the calibration of personal dosemeters at the Calibraion and Dosimetry laboratiry of the Peruvian Institute of Nuclear Energy (IPEN). Voltages applied to the X ray tube and the addition of filters have been made according to the publication ANSI No.13. Half value layer, homogeneity coefficient were found with respect to values published in ANSI No.13. Also spectral distributions and their resolutions were determined. Relative differences of 16 and 9% respectively were found in comparison with values calculated by the Montecarlo's Method.En el Laboratorio de Calibraciones y Dosimetría del IPEN se ha empleado el méodo de atenuación para caracterizar los haces de rayos X utilizados en la calibración de dosímetros personales. Los volatjes aplicados al tubo de rayos X y las filttraciones adicionales empleadas son sugeridos en el ANSI Nro. 13. A partir de los datos de atenuación se determinó la capa semirreductora, el coeficiente de homogeneidad y la energía media, encontrándose en promedio diferencias relativas de 3, 40 y 4% con respecto a los valores de ANSI Nro. 13. También se determinaron las distribuciones espectrales y sus resoluaciones; halándose diferencias relativas de 16 y 9% respectivamente, en comparación con los calculados por la técnica de Montecarlo.Universidad Nacional Mayor de San Marcos1999-04-02info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/859910.15381/rif.v1i02.8599Revista de Investigación de Física; Vol. 1 No. 02 (1998); 52-55Revista de Investigación de Física; Vol. 1 Núm. 02 (1998); 52-551728-29771605-7724reponame:Revistas - Universidad Nacional Mayor de San Marcosinstname:Universidad Nacional Mayor de San Marcosinstacron:UNMSMspahttps://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/8599/7445Derechos de autor 1999 T. Benavente, J. Márquez, J. Santa Cruzhttps://creativecommons.org/licenses/by-nc-sa/4.0info:eu-repo/semantics/openAccessoai:ojs.csi.unmsm:article/85992020-07-30T18:37:00Z
dc.title.none.fl_str_mv Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
Caracterización de Haces de Rayos X por el Método de Atenuación para la Calibración de Dosímetros Personales
title Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
spellingShingle Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
Benavente, T.
X ray
attenuation
spectra
dosimetry
radioprotection
rayos X
atenuación
espectros
dosímetría
radioprotección
title_short Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
title_full Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
title_fullStr Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
title_full_unstemmed Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
title_sort Characterization of X-ray beams by Attenuation Method for Calibration of Personal Dosimeters
dc.creator.none.fl_str_mv Benavente, T.
Márquez, J.
Santa Cruz, J.
author Benavente, T.
author_facet Benavente, T.
Márquez, J.
Santa Cruz, J.
author_role author
author2 Márquez, J.
Santa Cruz, J.
author2_role author
author
dc.subject.none.fl_str_mv X ray
attenuation
spectra
dosimetry
radioprotection
rayos X
atenuación
espectros
dosímetría
radioprotección
topic X ray
attenuation
spectra
dosimetry
radioprotection
rayos X
atenuación
espectros
dosímetría
radioprotección
description The choice of the attenuation method to characterize the X rays beams used in the calibration of personal dosemeters at the Calibraion and Dosimetry laboratiry of the Peruvian Institute of Nuclear Energy (IPEN). Voltages applied to the X ray tube and the addition of filters have been made according to the publication ANSI No.13. Half value layer, homogeneity coefficient were found with respect to values published in ANSI No.13. Also spectral distributions and their resolutions were determined. Relative differences of 16 and 9% respectively were found in comparison with values calculated by the Montecarlo's Method.
publishDate 1999
dc.date.none.fl_str_mv 1999-04-02
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv https://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/8599
10.15381/rif.v1i02.8599
url https://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/8599
identifier_str_mv 10.15381/rif.v1i02.8599
dc.language.none.fl_str_mv spa
language spa
dc.relation.none.fl_str_mv https://revistasinvestigacion.unmsm.edu.pe/index.php/fisica/article/view/8599/7445
dc.rights.none.fl_str_mv Derechos de autor 1999 T. Benavente, J. Márquez, J. Santa Cruz
https://creativecommons.org/licenses/by-nc-sa/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Derechos de autor 1999 T. Benavente, J. Márquez, J. Santa Cruz
https://creativecommons.org/licenses/by-nc-sa/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidad Nacional Mayor de San Marcos
publisher.none.fl_str_mv Universidad Nacional Mayor de San Marcos
dc.source.none.fl_str_mv Revista de Investigación de Física; Vol. 1 No. 02 (1998); 52-55
Revista de Investigación de Física; Vol. 1 Núm. 02 (1998); 52-55
1728-2977
1605-7724
reponame:Revistas - Universidad Nacional Mayor de San Marcos
instname:Universidad Nacional Mayor de San Marcos
instacron:UNMSM
instname_str Universidad Nacional Mayor de San Marcos
instacron_str UNMSM
institution UNMSM
reponame_str Revistas - Universidad Nacional Mayor de San Marcos
collection Revistas - Universidad Nacional Mayor de San Marcos
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1795238318175158272
score 13.871978
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).