Comparison between enzyme immunoassay and chemiluminescence methods for the quantification of thyroid hormones

Descripción del Articulo

Objective: To compare the sensitivity and specificity of enzyme immunoassay versus chemiluminescence for the quantification of thyroid hormones. Methods: This was a quantitative, observational, and cross-sectional study. The population consisted of 407 blood samples from patients seen at the outpati...

Descripción completa

Detalles Bibliográficos
Autor: Dávila-Colque, Isabel Paola
Formato: artículo
Fecha de Publicación:2025
Institución:Universidad de Huánuco
Repositorio:Revistas - Universidad de Huánuco
Lenguaje:español
OAI Identifier:oai:ojs2.localhost:article/752
Enlace del recurso:http://revistas.udh.edu.pe/index.php/RPCS/article/view/752
Nivel de acceso:acceso abierto
Materia:inmunoensayo
luminiscencia
hormonas tiroideas
sensibilidad y especificidad
immunoassay
luminescence
thyroid hormones
sensitivity and specificity
Descripción
Sumario:Objective: To compare the sensitivity and specificity of enzyme immunoassay versus chemiluminescence for the quantification of thyroid hormones. Methods: This was a quantitative, observational, and cross-sectional study. The population consisted of 407 blood samples from patients seen at the outpatient endocrinology clinic, with a final sample of 211 patients selected based on eligibility criteria. Thyroid hormone levels were measured using both enzyme immunoassay and chemiluminescence methods. A comparative analysis was performed to calculate sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). Results: The chemiluminescence method identified 68% of patients as negative and 32% as having thyroid disorders, all consistent with clinical fndings. The enzyme immunoassay reported 79% as negative and 21% as positive, of which 19% were true positives and 2% false positives. Additionally, 66% were true negatives and 13% false negatives. For the enzyme immunoassay, sensitivity was 57.9% and specificity 96.5%. The PPV was 88.9%, and the NPV was 83.6%. In contrast, the chemiluminescence method showed sensitivity, specificity, PPV, and NPV values exceeding 98%. The Kappa index was 0.6, indicating moderate agreement, where a value closer to 1 is ideal. Conclusions: The chemiluminescence method proved to be more sensitive and specific than enzyme immunoassay, making it the preferred method for quantifying thyroid hormones.
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).