Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms

Descripción del Articulo

The ultrasonic attenuation coefficient (ACE) can be used to classify tissue state. Pulse-echo spectral-based attenuation estimation techniques, such as the spectral-log-difference method (SLD), account for beam diffraction effects using a reference phantom having a sound speed close to the sound spe...

Descripción completa

Detalles Bibliográficos
Autores: Coila, Andres, Oelze, Michael L.
Formato: artículo
Fecha de Publicación:2020
Institución:Consejo Nacional de Ciencia Tecnología e Innovación
Repositorio:CONCYTEC-Institucional
Lenguaje:inglés
OAI Identifier:oai:repositorio.concytec.gob.pe:20.500.12390/2835
Enlace del recurso:https://hdl.handle.net/20.500.12390/2835
https://doi.org/10.1121/10.0001690
Nivel de acceso:acceso abierto
Materia:Arts and Humanities (miscellaneous)
Acoustics and Ultrasonics
http://purl.org/pe-repo/ocde/ford#2.11.02
id CONC_2fd51577f9b3ad77cace2385ebdfa2de
oai_identifier_str oai:repositorio.concytec.gob.pe:20.500.12390/2835
network_acronym_str CONC
network_name_str CONCYTEC-Institucional
repository_id_str 4689
dc.title.none.fl_str_mv Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
title Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
spellingShingle Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
Coila, Andres
Arts and Humanities (miscellaneous)
Acoustics and Ultrasonics
http://purl.org/pe-repo/ocde/ford#2.11.02
title_short Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
title_full Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
title_fullStr Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
title_full_unstemmed Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
title_sort Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms
author Coila, Andres
author_facet Coila, Andres
Oelze, Michael L.
author_role author
author2 Oelze, Michael L.
author2_role author
dc.contributor.author.fl_str_mv Coila, Andres
Oelze, Michael L.
dc.subject.none.fl_str_mv Arts and Humanities (miscellaneous)
topic Arts and Humanities (miscellaneous)
Acoustics and Ultrasonics
http://purl.org/pe-repo/ocde/ford#2.11.02
dc.subject.es_PE.fl_str_mv Acoustics and Ultrasonics
dc.subject.ocde.none.fl_str_mv http://purl.org/pe-repo/ocde/ford#2.11.02
description The ultrasonic attenuation coefficient (ACE) can be used to classify tissue state. Pulse-echo spectral-based attenuation estimation techniques, such as the spectral-log-difference method (SLD), account for beam diffraction effects using a reference phantom having a sound speed close to the sound speed of the sample. Methods like SLD assume linear propagation of ultrasound and do not account for potential acoustic nonlinear distortion of the backscattered power spectra in both sample and reference. In this study, the ACE of a sample was computed and compared using the SLD with two independent references (high attenuating and low attenuating phantoms but with similar B/A values) and over several pressure levels. Both numerical and physical tissue-mimicking phantoms were used in the study. The results indicated that the biases in ACE increased when using a reference having low attenuation, whereas the high attenuating reference produced more consistent ACE. Furthermore, increments in ACE vs input pressure were correlated to the log-ratio of Gol'dberg numbers between the sample and reference (R-2 = 0.979 in simulations and R-2 = 0.734 in experiments). Therefore, the results suggest that to reduce bias in ACE using spectral-based methods, both the sound speed and the Gol'dberg number of the reference phantom should be matched to the sample. (C) 2020 Acoustical Society of America.
publishDate 2020
dc.date.accessioned.none.fl_str_mv 2024-05-30T23:13:38Z
dc.date.available.none.fl_str_mv 2024-05-30T23:13:38Z
dc.date.issued.fl_str_mv 2020
dc.type.none.fl_str_mv info:eu-repo/semantics/article
format article
dc.identifier.uri.none.fl_str_mv https://hdl.handle.net/20.500.12390/2835
dc.identifier.doi.none.fl_str_mv https://doi.org/10.1121/10.0001690
url https://hdl.handle.net/20.500.12390/2835
https://doi.org/10.1121/10.0001690
dc.language.iso.none.fl_str_mv eng
language eng
dc.relation.ispartof.none.fl_str_mv JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv Acoustical Society of America (ASA)
publisher.none.fl_str_mv Acoustical Society of America (ASA)
dc.source.none.fl_str_mv reponame:CONCYTEC-Institucional
instname:Consejo Nacional de Ciencia Tecnología e Innovación
instacron:CONCYTEC
instname_str Consejo Nacional de Ciencia Tecnología e Innovación
instacron_str CONCYTEC
institution CONCYTEC
reponame_str CONCYTEC-Institucional
collection CONCYTEC-Institucional
repository.name.fl_str_mv Repositorio Institucional CONCYTEC
repository.mail.fl_str_mv repositorio@concytec.gob.pe
_version_ 1844883014500220928
spelling Publicationrp07702600rp07701600Coila, AndresOelze, Michael L.2024-05-30T23:13:38Z2024-05-30T23:13:38Z2020https://hdl.handle.net/20.500.12390/2835https://doi.org/10.1121/10.0001690The ultrasonic attenuation coefficient (ACE) can be used to classify tissue state. Pulse-echo spectral-based attenuation estimation techniques, such as the spectral-log-difference method (SLD), account for beam diffraction effects using a reference phantom having a sound speed close to the sound speed of the sample. Methods like SLD assume linear propagation of ultrasound and do not account for potential acoustic nonlinear distortion of the backscattered power spectra in both sample and reference. In this study, the ACE of a sample was computed and compared using the SLD with two independent references (high attenuating and low attenuating phantoms but with similar B/A values) and over several pressure levels. Both numerical and physical tissue-mimicking phantoms were used in the study. The results indicated that the biases in ACE increased when using a reference having low attenuation, whereas the high attenuating reference produced more consistent ACE. Furthermore, increments in ACE vs input pressure were correlated to the log-ratio of Gol'dberg numbers between the sample and reference (R-2 = 0.979 in simulations and R-2 = 0.734 in experiments). Therefore, the results suggest that to reduce bias in ACE using spectral-based methods, both the sound speed and the Gol'dberg number of the reference phantom should be matched to the sample. (C) 2020 Acoustical Society of America.Fondo Nacional de Desarrollo Científico y Tecnológico - FondecytengAcoustical Society of America (ASA)JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICAinfo:eu-repo/semantics/openAccessArts and Humanities (miscellaneous)Acoustics and Ultrasonics-1http://purl.org/pe-repo/ocde/ford#2.11.02-1Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantomsinfo:eu-repo/semantics/articlereponame:CONCYTEC-Institucionalinstname:Consejo Nacional de Ciencia Tecnología e Innovacióninstacron:CONCYTEC#PLACEHOLDER_PARENT_METADATA_VALUE##PLACEHOLDER_PARENT_METADATA_VALUE#20.500.12390/2835oai:repositorio.concytec.gob.pe:20.500.12390/28352024-05-30 15:25:41.672http://purl.org/coar/access_right/c_14cbinfo:eu-repo/semantics/closedAccessmetadata only accesshttps://repositorio.concytec.gob.peRepositorio Institucional CONCYTECrepositorio@concytec.gob.pe#PLACEHOLDER_PARENT_METADATA_VALUE##PLACEHOLDER_PARENT_METADATA_VALUE#<Publication xmlns="https://www.openaire.eu/cerif-profile/1.1/" id="acdad037-b09c-4862-8a15-7f508d018d3f"> <Type xmlns="https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types">http://purl.org/coar/resource_type/c_1843</Type> <Language>eng</Language> <Title>Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms</Title> <PublishedIn> <Publication> <Title>JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA</Title> </Publication> </PublishedIn> <PublicationDate>2020</PublicationDate> <DOI>https://doi.org/10.1121/10.0001690</DOI> <Authors> <Author> <DisplayName>Coila, Andres</DisplayName> <Person id="rp07702" /> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> <Author> <DisplayName>Oelze, Michael L.</DisplayName> <Person id="rp07701" /> <Affiliation> <OrgUnit> </OrgUnit> </Affiliation> </Author> </Authors> <Editors> </Editors> <Publishers> <Publisher> <DisplayName>Acoustical Society of America (ASA)</DisplayName> <OrgUnit /> </Publisher> </Publishers> <Keyword>Arts and Humanities (miscellaneous)</Keyword> <Keyword>Acoustics and Ultrasonics</Keyword> <Abstract>The ultrasonic attenuation coefficient (ACE) can be used to classify tissue state. Pulse-echo spectral-based attenuation estimation techniques, such as the spectral-log-difference method (SLD), account for beam diffraction effects using a reference phantom having a sound speed close to the sound speed of the sample. Methods like SLD assume linear propagation of ultrasound and do not account for potential acoustic nonlinear distortion of the backscattered power spectra in both sample and reference. In this study, the ACE of a sample was computed and compared using the SLD with two independent references (high attenuating and low attenuating phantoms but with similar B/A values) and over several pressure levels. Both numerical and physical tissue-mimicking phantoms were used in the study. The results indicated that the biases in ACE increased when using a reference having low attenuation, whereas the high attenuating reference produced more consistent ACE. Furthermore, increments in ACE vs input pressure were correlated to the log-ratio of Gol&apos;dberg numbers between the sample and reference (R-2 = 0.979 in simulations and R-2 = 0.734 in experiments). Therefore, the results suggest that to reduce bias in ACE using spectral-based methods, both the sound speed and the Gol&apos;dberg number of the reference phantom should be matched to the sample. (C) 2020 Acoustical Society of America.</Abstract> <Access xmlns="http://purl.org/coar/access_right" > </Access> </Publication> -1
score 13.425424
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).