Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras

Descripción del Articulo

Proper operation of the hardware and software of a SPECT Gamma Camera depends on its correct configuration and quality control. These are essential to obtain clinically valid images. In spite of the fact that these are recommended by the equipment manufacturer, these operations are not described in...

Descripción completa

Detalles Bibliográficos
Autores: Velásquez, María, Mantilla, Irla, Martí, Josep, Pachas, Alex
Formato: artículo
Fecha de Publicación:2017
Institución:Universidad Nacional de Ingeniería
Repositorio:Revista UNI - Tecnia
Lenguaje:español
OAI Identifier:oai:oai:revistas.uni.edu.pe:article/169
Enlace del recurso:http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169
Nivel de acceso:acceso abierto
Materia:equipos SPECT
líquidos miscibles
mezcla newtoniana
uniformidad extrínseca
uniformidad tomo gráfica
evaluación de operación total
id 0375-7765_1f19a9edefffd97cd22fbf8acdbaf57b
oai_identifier_str oai:oai:revistas.uni.edu.pe:article/169
network_acronym_str 0375-7765
repository_id_str .
network_name_str Revista UNI - Tecnia
dc.title.none.fl_str_mv Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
Construcción de un prototipo de mezcla bifásica radioactiva para evaluación de mapas de corrección intrínsecos en cámaras gamma Spect
title Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
spellingShingle Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
Velásquez, María
equipos SPECT
líquidos miscibles
mezcla newtoniana
uniformidad extrínseca
uniformidad tomo gráfica
evaluación de operación total
title_short Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
title_full Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
title_fullStr Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
title_full_unstemmed Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
title_sort Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma cameras
dc.creator.none.fl_str_mv Velásquez, María
Mantilla, Irla
Martí, Josep
Pachas, Alex
author Velásquez, María
author_facet Velásquez, María
Mantilla, Irla
Martí, Josep
Pachas, Alex
author_role author
author2 Mantilla, Irla
Martí, Josep
Pachas, Alex
author2_role author
author
author
dc.subject.none.fl_str_mv equipos SPECT
líquidos miscibles
mezcla newtoniana
uniformidad extrínseca
uniformidad tomo gráfica
evaluación de operación total
topic equipos SPECT
líquidos miscibles
mezcla newtoniana
uniformidad extrínseca
uniformidad tomo gráfica
evaluación de operación total
dc.description.none.fl_txt_mv Proper operation of the hardware and software of a SPECT Gamma Camera depends on its correct configuration and quality control. These are essential to obtain clinically valid images. In spite of the fact that these are recommended by the equipment manufacturer, these operations are not described in the manuals. The objective of the present work is to propose a unique method for the extrinsic quality control of any SPECT Gamma Camera manufactured to NEMA – IEC specifications. For this purpose, an automatic mixing module has been built to prepare uniform field phantoms, so as to carry out the quality control of the SPECT equipment. The module consists of an automatic mixer and a dummy (phantom). Thanks to the homogeneity of two‐ phase mixtures, we can obtain extrinsic correction maps which will serve to evaluate the quality of the equipment. This operation depends upon the homogeneity of the mixture and its optimal distribution in the phantom and in precise quantities, such as: sodium chloride 0.9% (3.5 liters) and sodium pertechnetate Tc 99m (?20mCi) from 0.1 to 0.4 ml. Radiation exposure must also be considered while preparing the mixture and trying to obtain an adequate dilution of a small amount of Tc‐99m in the volume of distilled water contained in the dummy in the shortest possible time. Analysis of the performance of the mixing module has been a very important research work. This depends upon an appropriate design of the mixing system: geometry of the impeller, mathematical modelling of the mixture, measurement of concentrations, various tests (geometry of the container, volume, diameter), speed and power of the motor, shape of the mixing blades (angle and number), all being important components to reach our objectives. The efficiency of the two‐phase mixture has been demonstrated when the SPECT Gamma Camera currently used by the INEN Nuclear Medicine Center obtained images of optimal quality.  
El buen funcionamiento del Hardware y Software de una Cámara Gamma SPECT, depende de la correcta configuración y del control de calidad. Estas acciones son esenciales para la obtención de imágenes clínicamente válidas. Los fabricantes, recomiendan estas acciones, sin embargo, no existen técnicas descritas en los manuales correspondientes a estos equipos. Los objetivos del presente trabajo consisten en proponer una técnica exclusiva para el control de calidad extrínseca de una Cámara Gamma SPECT de fabricación variable que estén reconocidas en NEMA, IEC. Para ello se construye un módulo mezclador automático para preparar fantomas de campo uniforme, mediante el cual nos permitirá realizar los controles de calidad de los equipos SPECT. El módulo está constituido por un equipo mezclador automático y de un maniquí (denominado fantoma). Mediante la homogeneidad de mezclas bifásicas, obtenemos mapas de corrección extrínsecos, las mismas que nos sirven para evaluar la calidad del equipo. Se puede observar que depende de la homogeneidad de la mezcla y su distribución óptima de la mezcla miscible en el fantoma y en cantidades de volúmenes precisos, tales como: Cloruro de sodio 0.9% (3.5 litros) y Pertecnetato de sodio Tc 99m (?20mCi) de 0.1 a 0.4 ml. Considerando además la exposición a la radiación al preparar la mezcla fluida al tratar de conseguir que una pequeña cantidad de Tc‐99m, se concentre adecuadamente por el volumen de agua destilada en el maniquí, en el menor tiempo posible. El rendimiento del módulo mezclador analizado ha sido un trabajo de investigación muy importante, éste depende del diseño apropiado del sistema mezclador (impulsor), conformado por la geometría, el modelo matemático de la mezcla, análisis de la concentración, pruebas (geometría recipiente, volumen, diámetros), velocidad (potencia del motor), forma de las paletas mezcladoras (ángulo y número) componentes muy importantes en los objetivos trazados. La eficacia de la mezcla bifásica se muestra con la adquisición óptima de las imágenes de calidad de la Cámara Gamma SPECT, que actualmente se usan en los centros de Medicina Nuclear, INEN.  
description Proper operation of the hardware and software of a SPECT Gamma Camera depends on its correct configuration and quality control. These are essential to obtain clinically valid images. In spite of the fact that these are recommended by the equipment manufacturer, these operations are not described in the manuals. The objective of the present work is to propose a unique method for the extrinsic quality control of any SPECT Gamma Camera manufactured to NEMA – IEC specifications. For this purpose, an automatic mixing module has been built to prepare uniform field phantoms, so as to carry out the quality control of the SPECT equipment. The module consists of an automatic mixer and a dummy (phantom). Thanks to the homogeneity of two‐ phase mixtures, we can obtain extrinsic correction maps which will serve to evaluate the quality of the equipment. This operation depends upon the homogeneity of the mixture and its optimal distribution in the phantom and in precise quantities, such as: sodium chloride 0.9% (3.5 liters) and sodium pertechnetate Tc 99m (?20mCi) from 0.1 to 0.4 ml. Radiation exposure must also be considered while preparing the mixture and trying to obtain an adequate dilution of a small amount of Tc‐99m in the volume of distilled water contained in the dummy in the shortest possible time. Analysis of the performance of the mixing module has been a very important research work. This depends upon an appropriate design of the mixing system: geometry of the impeller, mathematical modelling of the mixture, measurement of concentrations, various tests (geometry of the container, volume, diameter), speed and power of the motor, shape of the mixing blades (angle and number), all being important components to reach our objectives. The efficiency of the two‐phase mixture has been demonstrated when the SPECT Gamma Camera currently used by the INEN Nuclear Medicine Center obtained images of optimal quality.  
publishDate 2017
dc.date.none.fl_str_mv 2017-12-01
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Artículo evaluado por pares
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169
10.21754/tecnia.v27i2.169
url http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169
identifier_str_mv 10.21754/tecnia.v27i2.169
dc.language.none.fl_str_mv spa
language spa
dc.relation.none.fl_str_mv http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/44
http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/447
http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/536
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
audio/mpeg
text/html
dc.publisher.none.fl_str_mv Universidad Nacional de Ingeniería
publisher.none.fl_str_mv Universidad Nacional de Ingeniería
dc.source.none.fl_str_mv TECNIA; Vol 27 No 2 (2017); 7
TECNIA; Vol. 27 Núm. 2 (2017); 7
2309-0413
0375-7765
reponame:Revista UNI - Tecnia
instname:Universidad Nacional de Ingeniería
instacron:UNI
reponame_str Revista UNI - Tecnia
collection Revista UNI - Tecnia
instname_str Universidad Nacional de Ingeniería
instacron_str UNI
institution UNI
repository.name.fl_str_mv -
repository.mail.fl_str_mv mail@mail.com
_version_ 1701108799465586688
spelling Building of a radioactive two‐phase mixing prototype for the evaluation of extrinsic correction maps in spect gamma camerasConstrucción de un prototipo de mezcla bifásica radioactiva para evaluación de mapas de corrección intrínsecos en cámaras gamma SpectVelásquez, MaríaMantilla, IrlaMartí, JosepPachas, Alexequipos SPECTlíquidos misciblesmezcla newtonianauniformidad extrínsecauniformidad tomo gráficaevaluación de operación totalProper operation of the hardware and software of a SPECT Gamma Camera depends on its correct configuration and quality control. These are essential to obtain clinically valid images. In spite of the fact that these are recommended by the equipment manufacturer, these operations are not described in the manuals. The objective of the present work is to propose a unique method for the extrinsic quality control of any SPECT Gamma Camera manufactured to NEMA – IEC specifications. For this purpose, an automatic mixing module has been built to prepare uniform field phantoms, so as to carry out the quality control of the SPECT equipment. The module consists of an automatic mixer and a dummy (phantom). Thanks to the homogeneity of two‐ phase mixtures, we can obtain extrinsic correction maps which will serve to evaluate the quality of the equipment. This operation depends upon the homogeneity of the mixture and its optimal distribution in the phantom and in precise quantities, such as: sodium chloride 0.9% (3.5 liters) and sodium pertechnetate Tc 99m (?20mCi) from 0.1 to 0.4 ml. Radiation exposure must also be considered while preparing the mixture and trying to obtain an adequate dilution of a small amount of Tc‐99m in the volume of distilled water contained in the dummy in the shortest possible time. Analysis of the performance of the mixing module has been a very important research work. This depends upon an appropriate design of the mixing system: geometry of the impeller, mathematical modelling of the mixture, measurement of concentrations, various tests (geometry of the container, volume, diameter), speed and power of the motor, shape of the mixing blades (angle and number), all being important components to reach our objectives. The efficiency of the two‐phase mixture has been demonstrated when the SPECT Gamma Camera currently used by the INEN Nuclear Medicine Center obtained images of optimal quality.  El buen funcionamiento del Hardware y Software de una Cámara Gamma SPECT, depende de la correcta configuración y del control de calidad. Estas acciones son esenciales para la obtención de imágenes clínicamente válidas. Los fabricantes, recomiendan estas acciones, sin embargo, no existen técnicas descritas en los manuales correspondientes a estos equipos. Los objetivos del presente trabajo consisten en proponer una técnica exclusiva para el control de calidad extrínseca de una Cámara Gamma SPECT de fabricación variable que estén reconocidas en NEMA, IEC. Para ello se construye un módulo mezclador automático para preparar fantomas de campo uniforme, mediante el cual nos permitirá realizar los controles de calidad de los equipos SPECT. El módulo está constituido por un equipo mezclador automático y de un maniquí (denominado fantoma). Mediante la homogeneidad de mezclas bifásicas, obtenemos mapas de corrección extrínsecos, las mismas que nos sirven para evaluar la calidad del equipo. Se puede observar que depende de la homogeneidad de la mezcla y su distribución óptima de la mezcla miscible en el fantoma y en cantidades de volúmenes precisos, tales como: Cloruro de sodio 0.9% (3.5 litros) y Pertecnetato de sodio Tc 99m (?20mCi) de 0.1 a 0.4 ml. Considerando además la exposición a la radiación al preparar la mezcla fluida al tratar de conseguir que una pequeña cantidad de Tc‐99m, se concentre adecuadamente por el volumen de agua destilada en el maniquí, en el menor tiempo posible. El rendimiento del módulo mezclador analizado ha sido un trabajo de investigación muy importante, éste depende del diseño apropiado del sistema mezclador (impulsor), conformado por la geometría, el modelo matemático de la mezcla, análisis de la concentración, pruebas (geometría recipiente, volumen, diámetros), velocidad (potencia del motor), forma de las paletas mezcladoras (ángulo y número) componentes muy importantes en los objetivos trazados. La eficacia de la mezcla bifásica se muestra con la adquisición óptima de las imágenes de calidad de la Cámara Gamma SPECT, que actualmente se usan en los centros de Medicina Nuclear, INEN.  Universidad Nacional de Ingeniería2017-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionArtículo evaluado por paresapplication/pdfaudio/mpegtext/htmlhttp://www.revistas.uni.edu.pe/index.php/tecnia/article/view/16910.21754/tecnia.v27i2.169TECNIA; Vol 27 No 2 (2017); 7TECNIA; Vol. 27 Núm. 2 (2017); 72309-04130375-7765reponame:Revista UNI - Tecniainstname:Universidad Nacional de Ingenieríainstacron:UNIspahttp://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/44http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/447http://www.revistas.uni.edu.pe/index.php/tecnia/article/view/169/536info:eu-repo/semantics/openAccess2021-05-29T15:55:25Zmail@mail.com -
score 13.93557
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).