Construction of radon chamber to expose active and passive detectors

Descripción del Articulo

In this research and development, we present the design and manufacture of a radon chamber (PUCP radon chamber), a necessary tool for the calibration of passive detectors, verification of the operation of active radon monitors as well as diffusion chamber calibration used in radon measurements in ai...

Descripción completa

Detalles Bibliográficos
Autor: Rojas Hancco, Jhonny Jonnatan
Formato: tesis doctoral
Fecha de Publicación:2022
Institución:Pontificia Universidad Católica del Perú
Repositorio:PUCP-Institucional
Lenguaje:español
OAI Identifier:oai:repositorio.pucp.edu.pe:20.500.14657/189497
Enlace del recurso:http://hdl.handle.net/20.500.12404/24277
Nivel de acceso:acceso abierto
Materia:Radón
Radioactividad
Radón--Mediciones
https://purl.org/pe-repo/ocde/ford#1.03.00
id RPUC_cf6859df3842c5b14e6cd9f4922030a1
oai_identifier_str oai:repositorio.pucp.edu.pe:20.500.14657/189497
network_acronym_str RPUC
network_name_str PUCP-Institucional
repository_id_str 2905
spelling Lopez Herrera, Maria ElenaRojas Hancco, Jhonny Jonnatan2023-02-14T16:38:56Z2023-02-14T16:38:56Z20222023-02-14http://hdl.handle.net/20.500.12404/24277In this research and development, we present the design and manufacture of a radon chamber (PUCP radon chamber), a necessary tool for the calibration of passive detectors, verification of the operation of active radon monitors as well as diffusion chamber calibration used in radon measurements in air, and soils. The first chapter is an introduction to describe radon gas and national levels of radon concentration given by many organizations. Parameters that influence the calibration factor of the LR 115 type 2 film detector are studied, such as the energy window, critical angle, and effective volumes. Those are strongly related to the etching processes and counting of tracks all seen from a semi-empirical approach studied in the second chapter. The third chapter presents a review of some radon chambers that have been reported in the literature, based on their size and mode of operation as well as the radon source they use. The design and construction of the radon chamber are presented, use of uranium ore (autunite) as a chamber source is also discussed. In chapter fourth, radon chamber characterization is presented through leakage lambda, homogeneity of radon concentration, regimes-operation modes, and the saturation concentrations that can be reached. Procedures and methodology used in this work are contained in the fifth chapter and also some uses and applications of the PUCP radon chamber are presented; the calibration of cylindrical metallic diffusion chamber based on CR-39 chips detectors taking into account overlapping effect; transmission factors of gaps and pinhole for the same diffusion chambers are determined; permeability of glass fiber filter for 222Rn is obtained after reach equilibrium through Ramachandran model and taking into account a partition function as the rate of track density. The results of this research have been published in indexed journals. Finally, the conclusion and recommendations that reflect the fulfillment aims of this thesis are presented.spaPontificia Universidad Católica del PerúPEinfo:eu-repo/semantics/openAccesshttp://creativecommons.org/licenses/by-nc-nd/2.5/pe/RadónRadioactividadRadón--Medicioneshttps://purl.org/pe-repo/ocde/ford#1.03.00Construction of radon chamber to expose active and passive detectorsinfo:eu-repo/semantics/doctoralThesisTesis de doctoradoreponame:PUCP-Institucionalinstname:Pontificia Universidad Católica del Perúinstacron:PUCPDoctor en FísicaDoctoradoPontificia Universidad Católica del Perú. Escuela de PosgradoFísica07820505https://orcid.org/0000-0002-5028-441445645996533018Palacios Fernandez, Daniel FranciscoLopez Herrera, Maria ElenaNikezic, DragoslavSajo Bohus, LaszloQuindos Poncela, Luis Santiagohttps://purl.org/pe-repo/renati/level#doctorhttps://purl.org/pe-repo/renati/type#tesis20.500.14657/189497oai:repositorio.pucp.edu.pe:20.500.14657/1894972024-06-10 09:27:45.568http://creativecommons.org/licenses/by-nc-nd/2.5/pe/info:eu-repo/semantics/openAccessmetadata.onlyhttps://repositorio.pucp.edu.peRepositorio Institucional de la PUCPrepositorio@pucp.pe
dc.title.es_ES.fl_str_mv Construction of radon chamber to expose active and passive detectors
title Construction of radon chamber to expose active and passive detectors
spellingShingle Construction of radon chamber to expose active and passive detectors
Rojas Hancco, Jhonny Jonnatan
Radón
Radioactividad
Radón--Mediciones
https://purl.org/pe-repo/ocde/ford#1.03.00
title_short Construction of radon chamber to expose active and passive detectors
title_full Construction of radon chamber to expose active and passive detectors
title_fullStr Construction of radon chamber to expose active and passive detectors
title_full_unstemmed Construction of radon chamber to expose active and passive detectors
title_sort Construction of radon chamber to expose active and passive detectors
author Rojas Hancco, Jhonny Jonnatan
author_facet Rojas Hancco, Jhonny Jonnatan
author_role author
dc.contributor.advisor.fl_str_mv Lopez Herrera, Maria Elena
dc.contributor.author.fl_str_mv Rojas Hancco, Jhonny Jonnatan
dc.subject.es_ES.fl_str_mv Radón
Radioactividad
Radón--Mediciones
topic Radón
Radioactividad
Radón--Mediciones
https://purl.org/pe-repo/ocde/ford#1.03.00
dc.subject.ocde.es_ES.fl_str_mv https://purl.org/pe-repo/ocde/ford#1.03.00
description In this research and development, we present the design and manufacture of a radon chamber (PUCP radon chamber), a necessary tool for the calibration of passive detectors, verification of the operation of active radon monitors as well as diffusion chamber calibration used in radon measurements in air, and soils. The first chapter is an introduction to describe radon gas and national levels of radon concentration given by many organizations. Parameters that influence the calibration factor of the LR 115 type 2 film detector are studied, such as the energy window, critical angle, and effective volumes. Those are strongly related to the etching processes and counting of tracks all seen from a semi-empirical approach studied in the second chapter. The third chapter presents a review of some radon chambers that have been reported in the literature, based on their size and mode of operation as well as the radon source they use. The design and construction of the radon chamber are presented, use of uranium ore (autunite) as a chamber source is also discussed. In chapter fourth, radon chamber characterization is presented through leakage lambda, homogeneity of radon concentration, regimes-operation modes, and the saturation concentrations that can be reached. Procedures and methodology used in this work are contained in the fifth chapter and also some uses and applications of the PUCP radon chamber are presented; the calibration of cylindrical metallic diffusion chamber based on CR-39 chips detectors taking into account overlapping effect; transmission factors of gaps and pinhole for the same diffusion chambers are determined; permeability of glass fiber filter for 222Rn is obtained after reach equilibrium through Ramachandran model and taking into account a partition function as the rate of track density. The results of this research have been published in indexed journals. Finally, the conclusion and recommendations that reflect the fulfillment aims of this thesis are presented.
publishDate 2022
dc.date.created.none.fl_str_mv 2022
dc.date.accessioned.none.fl_str_mv 2023-02-14T16:38:56Z
dc.date.available.none.fl_str_mv 2023-02-14T16:38:56Z
dc.date.issued.fl_str_mv 2023-02-14
dc.type.es_ES.fl_str_mv info:eu-repo/semantics/doctoralThesis
dc.type.other.none.fl_str_mv Tesis de doctorado
format doctoralThesis
dc.identifier.uri.none.fl_str_mv http://hdl.handle.net/20.500.12404/24277
url http://hdl.handle.net/20.500.12404/24277
dc.language.iso.es_ES.fl_str_mv spa
language spa
dc.rights.es_ES.fl_str_mv info:eu-repo/semantics/openAccess
dc.rights.uri.*.fl_str_mv http://creativecommons.org/licenses/by-nc-nd/2.5/pe/
eu_rights_str_mv openAccess
rights_invalid_str_mv http://creativecommons.org/licenses/by-nc-nd/2.5/pe/
dc.publisher.es_ES.fl_str_mv Pontificia Universidad Católica del Perú
dc.publisher.country.es_ES.fl_str_mv PE
dc.source.none.fl_str_mv reponame:PUCP-Institucional
instname:Pontificia Universidad Católica del Perú
instacron:PUCP
instname_str Pontificia Universidad Católica del Perú
instacron_str PUCP
institution PUCP
reponame_str PUCP-Institucional
collection PUCP-Institucional
repository.name.fl_str_mv Repositorio Institucional de la PUCP
repository.mail.fl_str_mv repositorio@pucp.pe
_version_ 1835638203586969600
score 13.814859
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).