Reliability of a gas analyzer to determine glycemia compared to the conventional laboratory method
Descripción del Articulo
        Objectives: To determine the glucose reliability results measured in a gas equipment as compared with a reference method commonly used in the laboratory. The Clinical and Laboratory Standards Institute (CLSI) guide EP9- A2 instructions were followed. Design: Descriptive study with non-probability sa...
              
            
    
                        | Autor: | |
|---|---|
| Formato: | artículo | 
| Fecha de Publicación: | 2014 | 
| Institución: | Universidad Nacional Mayor de San Marcos | 
| Repositorio: | Revistas - Universidad Nacional Mayor de San Marcos | 
| Lenguaje: | español | 
| OAI Identifier: | oai:ojs.csi.unmsm:article/9776 | 
| Enlace del recurso: | https://revistasinvestigacion.unmsm.edu.pe/index.php/anales/article/view/9776 | 
| Nivel de acceso: | acceso abierto | 
| Materia: | Keywords Reliability serum glucose Clinical and Laboratory Standards Institute Guide EP9- A2 systematic error. Veracidad glucemia guía EP9–A2 del Clinical and Laboratory Standards Institute error sistemático. | 
| Sumario: | Objectives: To determine the glucose reliability results measured in a gas equipment as compared with a reference method commonly used in the laboratory. The Clinical and Laboratory Standards Institute (CLSI) guide EP9- A2 instructions were followed. Design: Descriptive study with non-probability sampling. Setting: Hospital Edgardo Rebagliati Martins, EsSalud, Lima, Peru. Materials: Blood sample of 234 subjects from the emergency services and intensive care unit. Methods: Blood glucose was processed with the ADVIA1800 equipment and the ABL800 gasometer. Results of both analyzers were compared following the mentioned guide directives, the Bland-Altman plot analysis and the Lin’s concordance correlation coefficient (CCC) calculation. Main outcome measures: Serum glucose concentration. Results: Average blood glucose levels obtained were 1.6 mg/dL higher for ABL800 than for ADVIA1800. Both methods showed a high positive correlation (beta coefficient 0.9995 and 95%, 95%CI 0.9994 to 0.9996). Glucose results for the method studied were acceptable, as confirmed with the Bland-Altman statistical analysis(0.9995 CCC value, 95%CI 0.9993 to 0.9996). Conclusions: The ABL800 analyzer is suitable for blood glucose monitoring, presenting an excellent correlation with the reference laboratory method. | 
|---|
 Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).
    La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).
 
   
   
             
            