Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
Descripción del Articulo
We present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with cap...
Autores: | , |
---|---|
Formato: | artículo |
Fecha de Publicación: | 2022 |
Institución: | Instituto Nacional Cardiovascular |
Repositorio: | Archivos peruanos de cardiología y cirugía cardiovascular |
Lenguaje: | español |
OAI Identifier: | oai:ojs.apcyccv.org.pe:article/210 |
Enlace del recurso: | https://apcyccv.org.pe/index.php/apccc/article/view/210 |
Nivel de acceso: | acceso abierto |
Materia: | marcapasos Falla del electrodo Vena subclavia Vena Axilar Pacemaker Lead failure Subclavian vein Axillary vein |
id |
REVINCOR_cb4a8d1a18b790821d235c0e53d70140 |
---|---|
oai_identifier_str |
oai:ojs.apcyccv.org.pe:article/210 |
network_acronym_str |
REVINCOR |
network_name_str |
Archivos peruanos de cardiología y cirugía cardiovascular |
repository_id_str |
|
spelling |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case reportRotura de aislante de electrodo de marcapasos por atrapamiento subclavio. Reporte de casoMeza Aquino, Miguel AlfonsoChoy Quintanilla, Emilio PablomarcapasosFalla del electrodoVena subclaviaVena AxilarPacemakerLead failureSubclavian veinAxillary veinWe present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with capture failure, so the entire pacing system was programmed to be replaced. During the intervention, breakage of the electrode insulator was observed. The subclavian approach can cause damage to the electrode due to entrapment in the soft tissues it passes through before entering the venous system and iterative stretching. Insulator breakage may manifest as loss of capture and presence of noise on the electrocardiogram, drop in lead impedance, and suggestive radiographic images. The axillary approach is a safe and effective option and is therefore recommended for transvenous lead implantation.Se presenta el caso de un paciente varón de 82 años con antecedente de implante de marcapasos unicameral hace 7 años, por bloqueo auriculoventricular de tercer grado, que experimenta presíncopes. En la telemetría se encuentró alerta de posible daño de cable electrodo y entrada de ruido ventricular que coincide con falla de captura, por lo que se programó reemplazo de todo el sistema de estimulación. Durante la intervención se observó rotura del aislante del electrodo. El abordaje subclavio puede ocasionar daño en el electrodo por atrapamiento en los tejidos blandos que atraviesa antes de ingresar al sistema venoso y estiramiento iterativo. La rotura del aislante se puede manifestar como pérdida de captura y presencia de ruido en el electrocardiograma, caída de la impedancia del cable e imágenes radiográficas sugerentes. El abordaje axilar es una opción segura y eficaz, por lo que se recomienda para el implante de cables transvenosos.Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud2022-06-30info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-Review articleArtículo evaluado por paresapplication/pdfapplication/epub+ziphttps://apcyccv.org.pe/index.php/apccc/article/view/210Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 No. 2 (2022); 127-131Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 Núm. 2 (2022); 127-1312708-7212reponame:Archivos peruanos de cardiología y cirugía cardiovascularinstname:Instituto Nacional Cardiovascularinstacron:INCORspahttps://apcyccv.org.pe/index.php/apccc/article/view/210/347https://apcyccv.org.pe/index.php/apccc/article/view/210/369Derechos de autor 2022 La revista es titular de la primera publicación, luego el autor dando crédito a la primera publicación.https://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessoai:ojs.apcyccv.org.pe:article/2102023-06-09T14:56:33Z |
dc.title.none.fl_str_mv |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report Rotura de aislante de electrodo de marcapasos por atrapamiento subclavio. Reporte de caso |
title |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
spellingShingle |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report Meza Aquino, Miguel Alfonso marcapasos Falla del electrodo Vena subclavia Vena Axilar Pacemaker Lead failure Subclavian vein Axillary vein |
title_short |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
title_full |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
title_fullStr |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
title_full_unstemmed |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
title_sort |
Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report |
dc.creator.none.fl_str_mv |
Meza Aquino, Miguel Alfonso Choy Quintanilla, Emilio Pablo |
author |
Meza Aquino, Miguel Alfonso |
author_facet |
Meza Aquino, Miguel Alfonso Choy Quintanilla, Emilio Pablo |
author_role |
author |
author2 |
Choy Quintanilla, Emilio Pablo |
author2_role |
author |
dc.subject.none.fl_str_mv |
marcapasos Falla del electrodo Vena subclavia Vena Axilar Pacemaker Lead failure Subclavian vein Axillary vein |
topic |
marcapasos Falla del electrodo Vena subclavia Vena Axilar Pacemaker Lead failure Subclavian vein Axillary vein |
description |
We present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with capture failure, so the entire pacing system was programmed to be replaced. During the intervention, breakage of the electrode insulator was observed. The subclavian approach can cause damage to the electrode due to entrapment in the soft tissues it passes through before entering the venous system and iterative stretching. Insulator breakage may manifest as loss of capture and presence of noise on the electrocardiogram, drop in lead impedance, and suggestive radiographic images. The axillary approach is a safe and effective option and is therefore recommended for transvenous lead implantation. |
publishDate |
2022 |
dc.date.none.fl_str_mv |
2022-06-30 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion Peer-Review article Artículo evaluado por pares |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
https://apcyccv.org.pe/index.php/apccc/article/view/210 |
url |
https://apcyccv.org.pe/index.php/apccc/article/view/210 |
dc.language.none.fl_str_mv |
spa |
language |
spa |
dc.relation.none.fl_str_mv |
https://apcyccv.org.pe/index.php/apccc/article/view/210/347 https://apcyccv.org.pe/index.php/apccc/article/view/210/369 |
dc.rights.none.fl_str_mv |
https://creativecommons.org/licenses/by/4.0 info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by/4.0 |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf application/epub+zip |
dc.publisher.none.fl_str_mv |
Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud |
publisher.none.fl_str_mv |
Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud |
dc.source.none.fl_str_mv |
Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 No. 2 (2022); 127-131 Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 Núm. 2 (2022); 127-131 2708-7212 reponame:Archivos peruanos de cardiología y cirugía cardiovascular instname:Instituto Nacional Cardiovascular instacron:INCOR |
instname_str |
Instituto Nacional Cardiovascular |
instacron_str |
INCOR |
institution |
INCOR |
reponame_str |
Archivos peruanos de cardiología y cirugía cardiovascular |
collection |
Archivos peruanos de cardiología y cirugía cardiovascular |
repository.name.fl_str_mv |
|
repository.mail.fl_str_mv |
|
_version_ |
1793529932186386432 |
score |
13.971837 |
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).