Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report

Descripción del Articulo

We present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with cap...

Descripción completa

Detalles Bibliográficos
Autores: Meza Aquino, Miguel Alfonso, Choy Quintanilla, Emilio Pablo
Formato: artículo
Fecha de Publicación:2022
Institución:Instituto Nacional Cardiovascular
Repositorio:Archivos peruanos de cardiología y cirugía cardiovascular
Lenguaje:español
OAI Identifier:oai:ojs.apcyccv.org.pe:article/210
Enlace del recurso:https://apcyccv.org.pe/index.php/apccc/article/view/210
Nivel de acceso:acceso abierto
Materia:marcapasos
Falla del electrodo
Vena subclavia
Vena Axilar
Pacemaker
Lead failure
Subclavian vein
Axillary vein
id REVINCOR_cb4a8d1a18b790821d235c0e53d70140
oai_identifier_str oai:ojs.apcyccv.org.pe:article/210
network_acronym_str REVINCOR
network_name_str Archivos peruanos de cardiología y cirugía cardiovascular
repository_id_str
spelling Rupture of pacemaker electrode insulator due to subclavian entrapment. Case reportRotura de aislante de electrodo de marcapasos por atrapamiento subclavio. Reporte de casoMeza Aquino, Miguel AlfonsoChoy Quintanilla, Emilio PablomarcapasosFalla del electrodoVena subclaviaVena AxilarPacemakerLead failureSubclavian veinAxillary veinWe present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with capture failure, so the entire pacing system was programmed to be replaced. During the intervention, breakage of the electrode insulator was observed. The subclavian approach can cause damage to the electrode due to entrapment in the soft tissues it passes through before entering the venous system and iterative stretching. Insulator breakage may manifest as loss of capture and presence of noise on the electrocardiogram, drop in lead impedance, and suggestive radiographic images. The axillary approach is a safe and effective option and is therefore recommended for transvenous lead implantation.Se presenta el caso de un paciente varón de 82 años con antecedente de implante de marcapasos unicameral hace 7 años, por bloqueo auriculoventricular de tercer grado, que experimenta presíncopes. En la telemetría se encuentró alerta de posible daño de cable electrodo y entrada de ruido ventricular que coincide con falla de captura, por lo que se programó reemplazo de todo el sistema de estimulación. Durante la intervención se observó rotura del aislante del electrodo. El abordaje subclavio puede ocasionar daño en el electrodo por atrapamiento en los tejidos blandos que atraviesa antes de ingresar al sistema venoso y estiramiento iterativo. La rotura del aislante se puede manifestar como pérdida de captura y presencia de ruido en el electrocardiograma, caída de la impedancia del cable e imágenes radiográficas sugerentes. El abordaje axilar es una opción segura y eficaz, por lo que se recomienda para el implante de cables transvenosos.Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud2022-06-30info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-Review articleArtículo evaluado por paresapplication/pdfapplication/epub+ziphttps://apcyccv.org.pe/index.php/apccc/article/view/210Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 No. 2 (2022); 127-131Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 Núm. 2 (2022); 127-1312708-7212reponame:Archivos peruanos de cardiología y cirugía cardiovascularinstname:Instituto Nacional Cardiovascularinstacron:INCORspahttps://apcyccv.org.pe/index.php/apccc/article/view/210/347https://apcyccv.org.pe/index.php/apccc/article/view/210/369Derechos de autor 2022 La revista es titular de la primera publicación, luego el autor dando crédito a la primera publicación.https://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessoai:ojs.apcyccv.org.pe:article/2102023-06-09T14:56:33Z
dc.title.none.fl_str_mv Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
Rotura de aislante de electrodo de marcapasos por atrapamiento subclavio. Reporte de caso
title Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
spellingShingle Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
Meza Aquino, Miguel Alfonso
marcapasos
Falla del electrodo
Vena subclavia
Vena Axilar
Pacemaker
Lead failure
Subclavian vein
Axillary vein
title_short Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
title_full Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
title_fullStr Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
title_full_unstemmed Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
title_sort Rupture of pacemaker electrode insulator due to subclavian entrapment. Case report
dc.creator.none.fl_str_mv Meza Aquino, Miguel Alfonso
Choy Quintanilla, Emilio Pablo
author Meza Aquino, Miguel Alfonso
author_facet Meza Aquino, Miguel Alfonso
Choy Quintanilla, Emilio Pablo
author_role author
author2 Choy Quintanilla, Emilio Pablo
author2_role author
dc.subject.none.fl_str_mv marcapasos
Falla del electrodo
Vena subclavia
Vena Axilar
Pacemaker
Lead failure
Subclavian vein
Axillary vein
topic marcapasos
Falla del electrodo
Vena subclavia
Vena Axilar
Pacemaker
Lead failure
Subclavian vein
Axillary vein
description We present the case of an 82-year-old male patient with a history of single-chamber pacemaker implantation 7 years ago, due to third-degree atrioventricular block, who experienced presyncope. Telemetry showed a warning of possible electrode lead damage and ventricular noise input coinciding with capture failure, so the entire pacing system was programmed to be replaced. During the intervention, breakage of the electrode insulator was observed. The subclavian approach can cause damage to the electrode due to entrapment in the soft tissues it passes through before entering the venous system and iterative stretching. Insulator breakage may manifest as loss of capture and presence of noise on the electrocardiogram, drop in lead impedance, and suggestive radiographic images. The axillary approach is a safe and effective option and is therefore recommended for transvenous lead implantation.
publishDate 2022
dc.date.none.fl_str_mv 2022-06-30
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Peer-Review article
Artículo evaluado por pares
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv https://apcyccv.org.pe/index.php/apccc/article/view/210
url https://apcyccv.org.pe/index.php/apccc/article/view/210
dc.language.none.fl_str_mv spa
language spa
dc.relation.none.fl_str_mv https://apcyccv.org.pe/index.php/apccc/article/view/210/347
https://apcyccv.org.pe/index.php/apccc/article/view/210/369
dc.rights.none.fl_str_mv https://creativecommons.org/licenses/by/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
application/epub+zip
dc.publisher.none.fl_str_mv Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud
publisher.none.fl_str_mv Instituto Nacional Cardiovascular “Carlos Alberto Peschiera Carrillo” – INCOR, EsSalud
dc.source.none.fl_str_mv Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 No. 2 (2022); 127-131
Archivos Peruanos de Cardiología y Cirugía Cardiovascular; Vol. 3 Núm. 2 (2022); 127-131
2708-7212
reponame:Archivos peruanos de cardiología y cirugía cardiovascular
instname:Instituto Nacional Cardiovascular
instacron:INCOR
instname_str Instituto Nacional Cardiovascular
instacron_str INCOR
institution INCOR
reponame_str Archivos peruanos de cardiología y cirugía cardiovascular
collection Archivos peruanos de cardiología y cirugía cardiovascular
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1793529932186386432
score 13.971837
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).