DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON

Descripción del Articulo

lt has been designed and built a conductimeter that operates with alternating current of 700 mÅ, 7.85 V and frequency among 50 - 60 Hz. For the voltage measures and amperage, a digital multimeter CIE- 123 with 2000 mΩ. Two electrodes were used, one built with two bars of amorphous coal and another o...

Descripción completa

Detalles Bibliográficos
Autor: Cjuno H., Jesús A.
Formato: artículo
Fecha de Publicación:2005
Institución:Universidad Nacional Mayor de San Marcos
Repositorio:Revista UNMSM - Revista Peruana de Química e Ingeniería Química
Lenguaje:español
OAI Identifier:oai:ojs.csi.unmsm:article/5468
Enlace del recurso:https://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468
Nivel de acceso:acceso abierto
Materia:Conductimeter
electrodes
amorphous
recycled coal.
Conductímetro
electrodos
carbón amorfo
reciclado.
id 1609-7599_755cea02254bab96c35788c1d7f29291
oai_identifier_str oai:ojs.csi.unmsm:article/5468
network_acronym_str 1609-7599
repository_id_str .
network_name_str Revista UNMSM - Revista Peruana de Química e Ingeniería Química
spelling DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBONDISEÑO Y ARMADO DE UN CONDUCTÍMETRO DE BAJA FRECUENCIA CON ELECTRODOS DE CARBÓN AMORFOCjuno H., Jesús A.Conductimeterelectrodesamorphousrecycled coal.Conductímetroelectrodoscarbón amorforeciclado.lt has been designed and built a conductimeter that operates with alternating current of 700 mÅ, 7.85 V and frequency among 50 - 60 Hz. For the voltage measures and amperage, a digital multimeter CIE- 123 with 2000 mΩ. Two electrodes were used, one built with two bars of amorphous coal and another of platinum normalized. For comparative ends, a normalized conductimeter Hanna HI98129 has been used. The values of conductivity obtained according to the official protocols, they are compatible with the references them and the comparison apparatus.Se ha diseñado y armado un conduclímetro que opera con corriente alterna de 700 mÅ, 7.85 V y frecuencia entre 50 - 60 Hz. Para las medidas de voltaje y amperaje, se ha hecho uso de un multímetro digital CIE-123 con resistencia de 2000 mΩ. Se utilizaron dos electrodos, uno de platino platinizado normalizado y otro, construido con dos varillas de carbón amorfo. Para fines comparativos, ha sido útil el conductímetro normalizado de la marca Hanna HI98129. Los valores de conductividad obtenidos según los protocolos oficiales, son compatibles con los referenciales y el aparato de comparación.Universidad Nacional Mayor de San Marcos2005-12-30info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468Revista Peruana de Química e Ingeniería Química; Vol 8 No 2 (2005); 23-28Revista Peruana de Química e Ingeniería Química; Vol. 8 Núm. 2 (2005); 23-281609-75991726-2208reponame:Revista UNMSM - Revista Peruana de Química e Ingeniería Químicainstname:Universidad Nacional Mayor de San Marcosinstacron:UNMSMspahttps://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468/4706Derechos de autor 2005 Jesús A. Cjuno H.http://creativecommons.org/licenses/by-nc-sa/4.0info:eu-repo/semantics/openAccess2021-06-01T18:11:04Zmail@mail.com -
dc.title.none.fl_str_mv DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
DISEÑO Y ARMADO DE UN CONDUCTÍMETRO DE BAJA FRECUENCIA CON ELECTRODOS DE CARBÓN AMORFO
title DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
spellingShingle DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
Cjuno H., Jesús A.
Conductimeter
electrodes
amorphous
recycled coal.
Conductímetro
electrodos
carbón amorfo
reciclado.
title_short DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
title_full DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
title_fullStr DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
title_full_unstemmed DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
title_sort DESIGN AND ASSEMBLY OF A LOW FREQUENCY CONDUCTIVITY ELECTRODES AMORPHOUS CARBON
dc.creator.none.fl_str_mv Cjuno H., Jesús A.
author Cjuno H., Jesús A.
author_facet Cjuno H., Jesús A.
author_role author
dc.subject.none.fl_str_mv Conductimeter
electrodes
amorphous
recycled coal.
Conductímetro
electrodos
carbón amorfo
reciclado.
topic Conductimeter
electrodes
amorphous
recycled coal.
Conductímetro
electrodos
carbón amorfo
reciclado.
dc.description.none.fl_txt_mv lt has been designed and built a conductimeter that operates with alternating current of 700 mÅ, 7.85 V and frequency among 50 - 60 Hz. For the voltage measures and amperage, a digital multimeter CIE- 123 with 2000 mΩ. Two electrodes were used, one built with two bars of amorphous coal and another of platinum normalized. For comparative ends, a normalized conductimeter Hanna HI98129 has been used. The values of conductivity obtained according to the official protocols, they are compatible with the references them and the comparison apparatus.
Se ha diseñado y armado un conduclímetro que opera con corriente alterna de 700 mÅ, 7.85 V y frecuencia entre 50 - 60 Hz. Para las medidas de voltaje y amperaje, se ha hecho uso de un multímetro digital CIE-123 con resistencia de 2000 mΩ. Se utilizaron dos electrodos, uno de platino platinizado normalizado y otro, construido con dos varillas de carbón amorfo. Para fines comparativos, ha sido útil el conductímetro normalizado de la marca Hanna HI98129. Los valores de conductividad obtenidos según los protocolos oficiales, son compatibles con los referenciales y el aparato de comparación.
description lt has been designed and built a conductimeter that operates with alternating current of 700 mÅ, 7.85 V and frequency among 50 - 60 Hz. For the voltage measures and amperage, a digital multimeter CIE- 123 with 2000 mΩ. Two electrodes were used, one built with two bars of amorphous coal and another of platinum normalized. For comparative ends, a normalized conductimeter Hanna HI98129 has been used. The values of conductivity obtained according to the official protocols, they are compatible with the references them and the comparison apparatus.
publishDate 2005
dc.date.none.fl_str_mv 2005-12-30
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv https://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468
url https://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468
dc.language.none.fl_str_mv spa
language spa
dc.relation.none.fl_str_mv https://revistasinvestigacion.unmsm.edu.pe/index.php/quim/article/view/5468/4706
dc.rights.none.fl_str_mv Derechos de autor 2005 Jesús A. Cjuno H.
http://creativecommons.org/licenses/by-nc-sa/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Derechos de autor 2005 Jesús A. Cjuno H.
http://creativecommons.org/licenses/by-nc-sa/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidad Nacional Mayor de San Marcos
publisher.none.fl_str_mv Universidad Nacional Mayor de San Marcos
dc.source.none.fl_str_mv Revista Peruana de Química e Ingeniería Química; Vol 8 No 2 (2005); 23-28
Revista Peruana de Química e Ingeniería Química; Vol. 8 Núm. 2 (2005); 23-28
1609-7599
1726-2208
reponame:Revista UNMSM - Revista Peruana de Química e Ingeniería Química
instname:Universidad Nacional Mayor de San Marcos
instacron:UNMSM
reponame_str Revista UNMSM - Revista Peruana de Química e Ingeniería Química
collection Revista UNMSM - Revista Peruana de Química e Ingeniería Química
instname_str Universidad Nacional Mayor de San Marcos
instacron_str UNMSM
institution UNMSM
repository.name.fl_str_mv -
repository.mail.fl_str_mv mail@mail.com
_version_ 1701389116619358208
score 13.945474
Nota importante:
La información contenida en este registro es de entera responsabilidad de la institución que gestiona el repositorio institucional donde esta contenido este documento o set de datos. El CONCYTEC no se hace responsable por los contenidos (publicaciones y/o datos) accesibles a través del Repositorio Nacional Digital de Ciencia, Tecnología e Innovación de Acceso Abierto (ALICIA).